Abstract
ABSTRACTTwo iPhone-sized differential mobility analyzers (DMAs) in the parallel-plate configuration (i.e., mini-plate DMAs) were designed and their performance was calibrated in this study in order to gain the instructive knowledge for the future mini-plate DMA design and to have a well-calibrated mini-plate DMA for the ultrafine particle (UFP) sensor network. The performance of mini-plate DMAs was calibrated using the tandem DMA (TDMA) technique. The experimental transfer functions of prototypes at different particle sizes and under various combinational conditions of aerosol and sheath flow rates were derived from the TDMA data. It is concluded that mini-plate DMAs performed reasonably well for UFP sizing. It was also found that the sizing resolution of mini-plate DMAs is closer to the aerosol-to-sheath flow rate ratio when the percentage of aerosol slit opening in length was increased (relative to the width of aerosol classification zone). A new concept of “effective sheath flow rate” was introduced to...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.