Abstract

Described are two types of direct-detection flat-panel X-ray detectors utilizing amorphous selenium (a-Se) and cadmium telluride (CdTe). The a-Se detector is fabricated using direct deposition onto a thin film transistor (TFT) substrate, whereas the CdTe detector is fabricated using a novel hybrid method, in which CdTe is pre-deposited onto a glass substrate and then connected to a TFT substrate. The detector array format is 512 X 512 with a pixel pitch of 150 micrometer. The imaging properties of both detectors have been evaluated with respect to X-ray sensitivity, lag, spatial resolution, and detective quantum efficiency (DQE). The modulation transfer functions (MTFs) measured at 1 lp/mm were 0.96 for a- Se and 0.65 for CdTe. The imaging lags after 33 ms were about 4% for a-Se and 22% for CdTe. The DQE values measured at zero spatial frequency were 0.75 for a-Se and 0.22 for CdTe. The results indicate that the a-Se and CdTe detectors have high potential as new digital X-ray imaging devices for both radiography and fluoroscopy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.