Abstract

The results of calibration of proton, cesium, potassium and helium–cesium scalar magnetometers in the range 20–100 μT are discussed. The calibration made it possible to estimate errors and to apply corrections to reduce the uncertainty of measurements. Experimental data have been obtained using a standard calibration system designed in the Russian Mendeleyev Metrology Institute (VNIIM). The systematic uncertainty in the reproduction of the magnetic flux density is estimated to be 1.8 × 10 −7, the random standard deviation being 0.001 nT. The experimental data can be used for the proper selection of measuring instruments and prediction of their uncertainty when studying the Earth's magnetic field.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.