Abstract
A complex neutron spectrum generated along with a useful photon beam imposes an additional radiation protection risk around medical linear accelerators (linac). The thermal neutron component of this complex neutron spectrum formed during different photon modes of operation of Elekta Versa HD linac has been quantified using Indium foil activation technique. The thermal neutron fluence (Φ th ) at isocenter for 15 MV, 10 MV and 10 MV FFF beams was found to be 2.45 × 105, 4.35 × 104 and 3.2 × 104 neutrons cm−2 Gy−1, respectively. The analysis shows a reduction in the Φ th as the flattening filter is being taken out from the beam path. A negative correlation in Φ th with respect to field size has been observed with an average 18% reduction in Φ th per monitor units as field size changes from 10 cm × 10 cm to 40 cm × 40 cm. For particular field size and photon energy, Φ th was found to be uniform across the patient plane. From the measured gamma ray spectrum inside the treatment room six major isotopes have been identified which were 122Sb, 187W, 82Br, 56Mn, 24Na and 28Al.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.