Abstract

Detection of glucose in water solution for several different concentrations has been performed with the purpose to determine the sensitivity of Near Infrared Bloch Surface Waves (lambda = 1.55microm) upon refractive index variations of the outer medium. TE-polarized electromagnetic surface waves are excited by a prism on a silicon nitride multilayer, according to the Kretschmann configuration. The real-time reflectance changes induced by discrete variations in glucose concentration has been revealed and analyzed. Without using any particular averaging strategy during the measurements, we pushed the device detection limit down to a glucose concentration of 2.5mg/dL, corresponding to a minimum detectable refractive index variation of the water solution as low as 3.8.10(-6).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.