Abstract

A series of measurements were undertaken using both high sensitivity radiochromic film and new lithium fluoride thermoluminescent dosimeters in a liquid water medium to define the radial dose function of 90Sr/90Y beta emitting intravascular brachytherapy sources more accurately. These measurements of a single 5 French source pellet served to verify current Monte Carlo transport models and extrapolation chamber measurements of the radial dose function, thus providing the recommended independent published measurements for g(r) of these sources. A slight deviation in the published radial dose function at depth leads the authors to recommend that treatment planning be performed using updated g(r) values from current Monte Carlo transport models verified by measurements such as those shown in this investigation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.