Abstract

The paper describes the procedure for evaluation of hemolysis induced by the Sputnik implantable pediatric rotary blood pump. In a series of five experiments the normalized index of hemolysis (N.I.H) for the Sputnik pump was found to be 0.0099 ± 0.0015 g/100 L.