Abstract

A newly designed encapsulated 103Pd source has been introduced (BrachySeed-Pd-103, also named Model Pd-1, manufactured by DRAXIMAGE Inc. and distributed by Cytogen Corp.) for interstitial brachytherapy to provide more isotropic dose distributions. In this work, the dosimetric characteristics of the 103Pd source were measured with micro LiF TLD chips and dosimetry parameters were characterized based upon the American Association of Physicists in Medicine (AAPM) Task Group No. 43 formalism. The dose rate constant of the sources was determined to be 0.66 +/-0.05 cGy h(-1) U(-1). The radial dose function was measured and was found to be similar to that of the Theragenics Model 200 103Pd source. The anisotropy constant for the Model Pd-1 source was determined to be 1.03.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.