Abstract

The LUPIN is a Rem counter for neutron dosimetry in pulsed radiation fields, i.e., those fields whose intensity varies greatly on short timescales with respect to the characteristic time of the utilized detector. This work describes the characterization of the energy response of the instrument. The response function was calculated with the Monte Carlo code MCNP6, representing the geometry and material composition of the LUPIN and simulating an irradiation in expanded and aligned monoenergetic neutron fields. The calculated response was validated in the monoenergetic fields of the National Physical Laboratory. The agreement between the calculated and measured responses is satisfactory, with a maximum discrepancy of 5%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.