Abstract

In the diagnostic frequency range, nonlinear imaging has been shown to improve image contrast and decrease artefacts. The extension of these techniques to high-frequency imaging (>15 MHz) was investigated. The second harmonic beam at 40 MHz of a high-frequency focused transducer (aperture 6 mm, focal distance 10 mm, f-number 1.67) was measured experimentally in water, in transmission and pulse-echo, and compared with the fundamental beams at 20 MHz and 40 MHz. Measurements were performed at peak negative pressures of 0.8 to 4.7 MPa. Transmission measurements were performed with a custom hydrophone with a 25-μm spot size to limit beam averaging. Over the range of peak negative pressures, the transmitted harmonic (40 MHz) beam had an average lateral beam width (−3 dB) of 77 μm and an average depth-of-field of 0.93 mm, whereas the fundamental beam had a corresponding beam width of 137 μm and a depth-of-field of 1.59 mm. The harmonic beam showed a 3-dB decrease in side lobe levels. Preliminary second harmonic images of mouse tissue in vitro are presented and compared to fundamental imaging at 20 and 40 MHz. (E-mail: emmanuel.cherin@swchsc.on.ca)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.