Abstract

This work relates to the study and characterization of the response function of an X-ray spectrometry system. The intrinsic efficiency of a Si(Li) detector has been simulated with the Monte Carlo codes MCNP and GEANT4 in the photon energy range of 2.6–59.5 keV. After finding it necessary to take a radiograph of the detector inside its cryostat to learn the correct dimensions, agreement within 10% between the simulations and experimental measurements with several point-like sources and PIXE results was obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.