Abstract

The gain achieved in image quality by utilizing, in the image forming process, the time-of-flight information (TOF) of positron annihilation photons between their inception and detection was measured experimentally by means of a positron emission tomograph (PET)-Super PETT I. The measurements were carried out by imaging a 35 cm cylindrical uniform phantom containing different positron activity concentrations. The gain achieved through the incorporation of TOF information, defined as the ratio of variances in images reconstructed with and without TOF information, was found to be approximately 3 at the lowest activity concentration and 5-8 in the activity concentration range typically encountered in clinical studies especially in fast or dynamic studies. This increase in gain with activity was interpreted as resulting from the reduction of random coincidences when TOF information is used. Further image improvement is yielded by incorporating TOF information into the PET attenuation correction provided by the measurement of transmission of annihilation photons in the object imaged.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.