Abstract

Magnetically guided capsule endoscopy is a promising technology for clinical application. A platform that simulates the magnetic capsule endoscope system is built to study the deformation process of the colon when its lumen suffers local forces. Force-displacement curves of the porcine large intestine under various experiment conditions, including different loading positions (haustra or taeniae coli), loading directions, colon inner pressures and specimen lengths, were measured to analyze the mechanical behavior of the intact large intestine during interactions with magnetic capsule endoscopes. In the practical application of the magnetic capsule endoscope, these data are imperative to optimize the control scheme and reduce operation risks. Based on our experiments, the taeniae coli of the intact large intestine show higher linear stiffness than the haustra, and inflation reduces the linear stiffness of the colon. Magnetic capsule with small edge radii can more easily damage or even perforate the colon. Based on our test results, we suggest that the force applied to the colon should be limited to below 17[Formula: see text]N when the capsule is actuated forward along the colon and limited to below 10[Formula: see text]N when the capsule is vertical to the colon during lesion screening.