Abstract

This report is the third volume in a three volume set describing the MARK series of pulsed ionizing radiation detection systems. This volume describes the MARK-1A detection system, compares it with the MARK-1 system, and describes the experimental testing of the detection systems. Volume 1 of this set presents the technical specifications for the MARK-1 detection system. Volume 2 is an operations manual specifically for the MARK-1 system, but it generally applies to the MARK-1A system as well. These detection systems operate remotely and detect photon radiation from a single or a multiple pulsed source. They contain multiple detector (eight in the MARK-1 and ten in the MARK-1A) for determination of does and incident photon effective energy. The multiple detector arrangement, having different detector sizes and shield thicknesses, provides the capability of determining the effective photon energy of the radiation spectrum. Dose measurements using these units are consistent with TLD measurements. The detection range is from 3 nanorads to 90 microrads per source burst; the response is linear over that range. Three units were built and are ready for field deployment.