Abstract

BackgroundIn computed tomography (CT) quality assurance, alignment of image quality phantoms is crucial for quantitative and reproducible evaluation and may be improved by alignment correction. Our goal was to develop an alignment correction algorithm to facilitate geological sampling of sediment cores taken from a cold-water coral mount.MethodsAn alignment correction algorithm was developed and tested with a CT acquisition at 120 kVp and 150 mAs of an image quality phantom. Random translation (maximum 15 mm) and rotation (maximum 2.86°) were applied and ground-truth was compared to parameters determined by alignment correction. Furthermore, mean densities were evaluated in four regions of interest (ROIs) placed in the phantom low-contrast section, comparing values before and after correction to ground truth. This process was repeated 1000 times. After validation, alignment correction was applied to CT acquisitions (140 kVp, 570 mAs) of sediment core sections up to 1 m in length, and sagittal reconstructions were calculated for sampling planning.ResultsIn the phantom, average absolute differences between applied and detected parameters after alignment correction were 0.01 ± 0.06 mm (mean ± standard deviation) along the x-axis, 0.11 ± 0.08 mm along the y-axis, 0.15 ± 0.07° around the x-axis, and 0.02 ± 0.02° around the y-axis, respectively. For ROI analysis, differences in densities were 63.12 ± 30.57, 31.38 ± 32.10, 18.27 ± 35.57, and 9.59 ± 26.37 HU before alignment correction and 1.22 ± 1.40, 0.76 ± 0.9, 0.45 ± 0.86, and 0.36 ± 0.48 HU after alignment correction, respectively. For sediment core segments, average absolute detected parameters were 3.93 ± 2.89 mm, 7.21 ± 2.37 mm, 0.37 ± 0.33°, and 0.21 ± 0.22°, respectively.ConclusionsThe alignment correction algorithm was successfully evaluated in the phantom and allowed a correct alignment of sediment core segments, thus aiding in sampling planning. Application to other tasks, like image quality analysis, seems possible.

Highlights

  • In computed tomography (CT) quality assurance, alignment of image quality phantoms is crucial for quantitative and reproducible evaluation and may be improved by alignment correction

  • On board of the vessel, the sediment core was split into sections of 1 m, but remained unopened as a prerequisite for subsequent CT imaging at the Clinic of Diagnostic and Interventional Radiology (DIR) of Heidelberg University Hospital, Heidelberg, Germany

  • Sagittal images were reconstructed at a slice thickness of 5 mm with an increment of 5 mm, adding a digital ruler as an Average differences between generated parameters and the parameters detected by the algorithm, which can be used to investigate a bias in the difference, were -0.03 ± 0.09 mm for translation along the x-axis, 0.00 ± 0.16 mm for translation along the y-axis, 0.11 ± 0.11° for rotation around the x-axis, and 0.00 ± 0.05° for rotation around the y-axis

Read more

Summary

Introduction

In computed tomography (CT) quality assurance, alignment of image quality phantoms is crucial for quantitative and reproducible evaluation and may be improved by alignment correction. Our goal was to develop an alignment correction algorithm to facilitate geological sampling of sediment cores taken from a cold-water coral mount. Accurate positioning of scanned objects as well as of patients is important for ensuring adequate image quality and stability of measured computed tomography (CT) numbers. Quantitative evaluations, e.g., of standardised image quality phantoms, rely on the exact positioning of the imaged object to produce reproducible and accurate results [1]. The qualitative information on coral occurrences in the sediment core further helps for accurate sampling with minimal disturbance of sediment and other fragments by allowing to locate CWC fragments below the surface

Objectives
Methods
Results
Conclusion