Abstract
The effect of detector collimator and scatterer thickness on multiple Compton backscattered gamma photons is studied. Gamma photons from a 137Cs source of 5.8 mCi, is allowed to fall on Carbon, Aluminium, Iron and Copper targets and the scattered photons are detected by a properly shielded 76 mm × 76 mm NaI(Tl) scintillation detector located at 90° to the incident beam. To extract the contribution of multiple scattered photons from the measured spectra, single scattered distribution is remodelled analytically. The thickness at which the multiple scattered photons saturate is determined for different detector collimator apertures and scatterer thicknesses. The variation of saturation thickness, Signal-to-Noise (S/N) ratio, Multiple Scattering Fraction (MSF) for different materials and collimator sizes are studied and compared with the available literature. Monte Carlo simulated calculations using MCNP code supports the present experimental work.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Radiation Research and Applied Sciences
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.