Abstract
Abstract We describe the results of experimental studies of Liquid Encapsulated Czochralski (LEC) semi-insulating GaAs detectors with different contact areas. The detector active area spreading at different applied bias is proven by three various experiments. In the first, the scan of the detector electrode and adjacent area was performed using the 660 nm pulsed laser beam, with a spot diameter of 50 μm. The other experiments use the spectra measured by tested detectors, when irradiated by 241 Am α source (5.48 MeV). The experiments show relation between bias voltage applied and the range of active detector area from the edge of the detector contact.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.