Abstract

One of a critical problem in the biomedical signals measurements is hum noise activity due to power line interference. Various approaches to suppress hum noise both analog and digital techniques have been proposed. However, these approaches have some disadvantages. N-path notch filter can be an alternative solution for this problem. The notch depth in a conventional N-path notch filter is limited by the number of paths to achieve deeper notch. This paper proposes a new N-path notch filter with additional sample-and-hold (S/H) and leak buffer circuits to improve notch depth. Simulation and measurement results of 3-phase N-path filter achieve notch depth of 61.6dB and 54.5dB respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.