Scintillation detectors with light readout are used for gamma, x-ray and particle detection. Where applicable, photon counting is principally superior to charge integration with regard to accuracy. Most scintillation detectors, however, generate a large number of photons per microsecond for a typical scintillation pulse resulting in significant amounts of pileup. This precludes the separation, and thus direct counting of single photoelectron charges.The algorithm developed and presented in this paper quantifies the coarseness of fast digitized current tracks to construct a photon count dependent, however, electron gain independent charge calculation function. Underlying photoelectrons are interpreted as noise components and retrieved by a modified statistical variance calculation. This method is verified for modeled scintillation pulses and scintillation detector data. It provides a new means for PMT gain stabilization in digital multi-channel analyzers by pulse current analysis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.