Abstract
The purpose of this paper is to present a statistical reviewer's perspective on some technical aspects of reviewing Bayesian medical device trials submitted to the Food and Drug Administration. The discussion reflects the experiences of the authors and should not be misconstrued as official guidance by the FDA. A variety of applications are described, reflecting our experience with therapeutic and diagnostic devices. In addition to Bayesian analysis of trials, Bayesian trial design and Bayesian monitoring are discussed. Analyses were implemented in WinBUGS (http://www.mrc-bsu.cam.ac.uk/bugs/winbugs/contents.shtml), with the code provided.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.