Abstract

In an experiment to measure the D(t,?)n cross section at beam energies of 10 to 120 keV, we have developed a double-compensating beam calorimeter, based on a Swiss Design to measure the particle beam intensity. A Faraday cup is not useful because of considerable charge exchange in the target gas at such low beam energies. We calibrated the calorimeter both with 10- and 3-MeV protons (comparing with a Faraday-cup measurement of the beam flux) and with the heat generated in a precision resistor. Both methods agree and give a calibration accurate to ±0.08% over a range of 10 to 800 mW beam power. Beam powers as low as 5 mW may be used, but with less accuracy. The beam energy must be known in order to calculate the particle intensity. Some difficulties with and peculiarities of the device are discussed.