Abstract
We applied the heat citric acid disinfection method in the main part of the central dialysis fluid delivery system (MPCDDS), which consists of a multiple-patient dialysis fluid supply unit, dialysis console units, and dialysis fluid piping. This disinfection method has been used for single-patient dialysis machines, but this is the first trial in the MPCDDS. We examined, by points of safety and disinfection effect, whether this disinfection method is comparable to conventional disinfection methods in Japan. The conventional disinfection method is a combination of two disinfectants, sodium hypochlorite and acetic acid, used separately for protein removal and decalcification. Consequently, total microbial counts and endotoxin concentrations fully satisfied the microbiological requirements for standard dialysis fluid of ISO 11663. From our results and discussion, this heat citric acid disinfection method is proved to be safe and reliable for MPCDDS. However, to satisfy the microbiological requirements for ultrapure dialysis fluid, further consideration for this method in MPCDDS including the reverse osmosis device composition and piping is necessary.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.