Abstract

To describe an on-table modification of standard angiography catheters for use in directed arterial and venous thrombolysis. An angiogram is performed and the length of thrombosed vessel (artery or vein) is measured. A 5 or 6 Fr catheter (preferably straight/multi- purpose/vertebral catheter) is modified on table for use by making multiple holes with 23 G needle. After testing ex vivo with saline injection, the on table modified catheter is placed over a wire into the thrombosed segment of the vessel and thrombolytic agent infusion is commenced utilizing a syringe driver after giving a bolus dose of thrombolytic agent. Median duration of thrombolysis was 24h in our study. We have utilized this method in twenty thrombosed vessels, without any catheter related complications. In our experience, this modification of a standard catheter as a multi-hole catheter is a readily available, simple, cheap, versatile and effective device for directed thrombolysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.