Abstract
Rad 8 is a small, in-house computer system for calculating radiotherapy dose distributions, both for multiple field and rotational external beam techniques and for interstitial and intracavitary redium treatments. Implementing the system has its difficulties, particularly in obtaining the necessary beam profile information. Several approaches have been tried and finally a separate profile generator program was written. Other difficulties encountered with beam data storage were overcome by making minor changes to the system programs. The various calculation techniques have been evaluated, and in general found to be satisfactory, and suitable factors to be used in the inhomogeneity correction for 4 MV X rays have been calculated based on previously published measured data. An option to change beam weightings exists and this can be used to renormalize dose distributions to give percentage isodose curves relative to the prescribed tumour dose. If this procedure is followed dose homogeneity over the target volume can readily be assessed, and the exposure times may easily be calculated from the new weightings. Certain shortcomings in the rotating beam program and the radium program may easily lead to error if extreme care is not exercised.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.