Abstract

A strategy for expanding the linear working range in bioanalysis using quantitative high performance liquid chromatography/tandem mass spectrometry (HPLC/MS/MS) is presented. The strategy involves monitoring multiple product ions. Herein we demonstrate the strategy on a rat plasma assay for a proprietary experimental drug where the linear range is expanded from 2 to 4 orders of magnitude. A primary sensitive ion was monitored to obtain a high sensitivity range calibration curve (0.400 to 100 ng/mL) while a less sensitive secondary ion was monitored to obtain a low sensitivity range calibration curve (90.0 to 4000 ng/mL). Each calibration curve gave acceptable linearity (r >0.990). Quality control samples at low, mid and high levels within each calibration curve demonstrated acceptable precision and accuracy (within 20% for all levels). The technique was successfully applied to rat pre-clinical sample analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.