Abstract

Accurate and sensitive detection of ochratoxin A (OTA) is highly necessary due to its high carcinogenicity, teratogenicity and mutagenicity. Herein, we reported an exogenous interference and autofluorescence-free ratiometric aptasensor based on dual-colored persistent luminescent nanoparticles for precise detection of OTA. Green-emitting ZnGeO:Mn bonded with OTA aptamer and BHQ1-modified complementary base was acted as detection and specific recognition probe (ZGM@BHQ1). Quaternary ammonium modified ZnGaGeO:Cr with red emission was employed as reference probe and further bonded to ZGM@BHQ1 through electrostatic interaction to construct the ratiometric aptasensor. The developed ratiometric aptasensor was free from real-time excitation, external interference and autofluorescence and gave low detection limit of 3.4 pg mL−1, wide linearity in the range of 0.01–50 ng mL−1 and high precision of 3.1 % (11 replicate determinations, at 1 ng mL−1 level). The applicability of the aptasensor was successfully demonstrated by analyzing OTA in in grain samples with recoveries of 97.6 %-105.2 %.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call