Abstract

PurposeExercise cardiovascular magnetic resonance (Ex-CMR) typically requires complex post-processing or transient exercise cessation, decreasing clinical utility. We aimed to demonstrate the feasibility of assessing biventricular volumes and great vessel flow during continuous in-scanner Ex-CMR, using vendor provided Compressed SENSE (C-SENSE) sequences and commercial analysis software (Cvi42).Methods12 healthy volunteers (8-male, age: 35 ± 9 years) underwent continuous supine cycle ergometer (Lode-BV) Ex-CMR (1.5T Philips, Ingenia). Free-breathing, respiratory navigated C-SENSE short-axis cines and aortic/pulmonary phase contrast magnetic resonance (PCMR) sequences were validated against clinical sequences at rest and used during low and moderate intensity Ex-CMR. Optimal PCMR C-SENSE acceleration, C-SENSE-3 (CS3) vs C-SENSE-6 (CS6), was further investigated by image quality scoring. Intra-and inter-operator reproducibility of biventricular and flow indices was performed.ResultsAll CS3 PCMR image quality scores were superior (p < 0.05) to CS6 sequences, except pulmonary PCMR at moderate exercise. Resting stroke volumes from clinical PCMR sequences correlated stronger with CS3 than CS6 sequences. Resting biventricular volumes from CS3 and clinical sequences correlated very strongly (r > 0.93). During Ex-CMR, biventricular end-diastolic volumes (EDV) remained unchanged, except right-ventricular EDV decreasing at moderate exercise. Biventricular ejection-fractions increased at each stage. Exercise biventricular cine and PCMR stroke volumes correlated very strongly (r ≥ 0.9), demonstrating internal validity. Intra-observer reproducibility was excellent, co-efficient of variance (COV) < 10%. Inter-observer reproducibility was excellent, except for resting right-ventricular, and exercise bi-ventricular end-systolic volumes which were good (COV 10–20%).ConclusionBiventricular function, aortic and pulmonary flow assessment during continuous Ex-CMR using CS3 sequences is feasible, reproducible and analysable using commercially available software.

Highlights

  • Stress cardiac imaging is an important tool in assessing valvular [1] and congenital heart disease [2] and has significantly improved the diagnostic accuracy for coronary artery disease (CAD) detection compared to exercise ECG [3, 4]

  • Protocol development and feasibility testing was achieved by: (1) developing a free-breathing Compressed SENSE (C-SENSE) protocol and validating this against our institute’s standard clinical imaging sequences at rest; (2) determining the optimal acceleration of C-SENSE for phase contrast magnetic resonance (PCMR) sequences, for use in Exercise Cardiovascular magnetic resonance (CMR) (Ex-CMR), by assessing resting and exercise image quality and comparing the derived stroke volumes against standard clinical imaging sequences at rest; (3) utilising the validated C-SENSE protocol during continuous low and moderate exercise intensities to determine if the acquired biventricular volumes and flow have internal validity in terms of consistency of ventricular stroke volumes when derived separately from cavity volumes and great vessel flow measurements, and whether they are concordant with expected supine exercise physiology

  • Mean resting aortic and pulmonary stroke volumes acquired from all 4 PCMR sequences were comparable, with CS3 and CS6 free-breathing flow showing minimal bias with both breathhold and free-breathing standard clinical flow sequences (Table 2)

Read more

Summary

Introduction

Stress cardiac imaging is an important tool in assessing valvular [1] and congenital heart disease [2] and has significantly improved the diagnostic accuracy for coronary artery disease (CAD) detection compared to exercise ECG [3, 4]. The continued need for cardiac gating resulted in detrimental artefacts at maximal exercise and real time Ex-CMR studies assessing flow report the acquisition of a significant volume of flow data (< 25,000 images per patient), requiring the use of an online graphics processing unit reconstruction system and prolonged post processing/analysis time [17]. The development of ungated real-time cine imaging solved the ECG gating issues, allowing biventricular volume assessment during maximal exercise [14]. Combining this technique with ungated flow acquisition resulted in the first study assessing bi-ventricular volumes and aortic and pulmonary flow during continuous exercise [18]. The aims of this study are to demonstrate the feasibility of assessing biventricular volume and flow during continuous exercise using vendor provided C-SENSE sequences and commercially available standard analysis software

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.