Abstract

The present study reports time-and spatially-dependent emission intensities and corresponding excitation temperatures for a pulse-operated argon microwave-induced plasma. These measurements were made at a variety operating conditions. Excitation temperatures were determined using argon, copper, and iron as thermometric species. Observed temperatures show significant differences with respect to each other. Argon emission lines exhibit a pronounced afterglow enhancement which is not observed for analytical lines. The relation-ship between excitation temperature and afterglow enhancement was studied. Excitation temperature was found to decrease suddenly at the falling edge of the microwave pulse.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.