Abstract
The purpose of this study was to evaluate Interaural Attenuation (IA) in frequency base in the insert earphones that are used in audiological assessments. Thirty healthy subjects between 18-65 years of age (14 female and 16 male) participated in our study. Otoscopic examination was performed on all participants. Audiological evaluations were performed using the Interacoustics AC40 clinical audiometer and ER-3A insert earphones. IA value was calculated by subtracting good ear bone conduction hearing thresholds of the worst airway hearing threshold. In our measuring for 0.125-8.0 kHz frequency were performed in our audiometry device separately for each frequency. IA amount in the results we found in 1000 Hz and below frequencies about 75-110 dB range avarage is 89±5dB, in above 1000 Hz frequencies in 50-95 dB range and avarage it is changed to 69±5dB. According to the obtained findings the quantity of melting in the transition between the ears are increasing with the insert earphones. The insert earphone should be beside supraaural earphone that is routinely used in clinics. Difficult masking applications due to the increase in the value of IA can be easily done with insert earphones.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.