Abstract

A technique is described for examining the mode of operation of radiation detectors and estimating the losses and gains caused by random summing effects in spectra acquired at high counting rates. The technique is applied to estimating the errors caused by random summing when measuring the amounts of carbon and nitrogen in the human body by the prompt gamma ray neutron activation analysis system installed in Leeds, UK. The authors' examination shows that the counting rate would have to increase to more than twice the value used at the method's calibration, before an error of greater than 10% is caused in the estimation of carbon by random summing effects, whilst for nitrogen a change of only 18% in the counting rate will cause a 10% error in its estimation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.