Abstract

In this paper, newly application of the absorptiometry microchip was considered and sensor property was characterized with real sample. C-reactive protein (CRP) measurement, that is one of the items of serology test, was chosen as first application. CRP value is very useful marker of irritation and also appropriate to confirm ability of sensor performance for Particle-Enhanced Turbidimetric Immunoassay (PETIA). Fabricated microchip shows characteristics equivalent to commercially available sensor unit. Sensor property with different concentration samples was also evaluated. Temperature effect for PETIA was described with characterized data. From the results of characterization and discussion, the microchip can be applicable for CRP measurement with PETIA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.