Abstract

Dosing performance of dry powder inhalers is dependent on patient's inspiratory effort. This study compares the inhalation profiles generated by patients with severe obstructive lung disease using DiskusTMand TurbuhalerTMinhalers. The patient profiles are subsequently used to determine the dosing performance of fluticasone propionate DiskusTMand budesonide TurbuhalerTMinhalers.Inhalation profiles were recorded in COPD patients (FEV1≤30% predicted) as they inhaled with maximal effort through the inhalers. The profiles were used in an inhalation simulator to assess the dosing performance by measuring the total emitted dose and the fine particle mass for each inhaler type.Peak inspiratory flow was significantly higher through the DiskusTM(mean 82·3 l min−1) compared with TurbuhalerTM(mean 53·5 l min−1, difference = 28.8 l min−1;P<0·0001). In addition, in direct comparison of the two devices, the DiskusTMwas shown to deliver a more consistent dose irrespective of flow than the TurbuhalerTMin this patient population. These findings may be of importance in optimising selection of devices for patients with severe airway obstruction.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.