Abstract
The fast spin-echo (FSE) sequence is frequently used as a fast data-readout technique in conjunction with other pulse sequence elements, such as in fluid-attenuated inversion-recovery (FLAIR) and double inversion-recovery (DIR) sequences. In order to implement those pulse sequences, an understanding is required of how the longitudinal magnetization evolves during the FSE part of the sequence. This evolution has been addressed to a certain extent by previous publications, but the DIR literature in particular appears to be replete with approximations to the exact expression for the longitudinal magnetization, and several papers contain errors. Equations are therefore presented here for the evolution of the longitudinal magnetization for a FSE readout. These are then applied to calculate the magnetization available immediately prior to the 90 degrees imaging pulse for the FLAIR-FSE and DIR-FSE pulse sequences.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.