Abstract

The purpose of the present study was to investigate the retest reliability of event-related functional near-infrared spectroscopy (fNIRS). Therefore, isolated functional activation was evoked in the occipital cortex by a periodic checkerboard stimulation. During a 52-channel fNIRS recording, 12 subjects underwent 60 trials of visual stimulation in two sessions. The retest interval was set to 3 weeks. Linear correlations of the contrast t values supplemented by scatter plots, channel-wise intraclass correlation coefficients (ICC) as well as reproducibility indices for the quantity of activated channels ( R QUANTITY) and the location ( R OVERLAP) of the detected activation were calculated. The results at the group level showed good reliability in terms of the single measure ICCs (up to 0.84) and excellent reproducibility quantified by R QUANTITY and R OVERLAP (up to 96% of the quantity and the location were reproducible), whereas the results at the single subjects' level were mediocre. Furthermore, the reliability assessed by single measurement ICCs improved if regarded at a cluster level.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call