Abstract

Abstract We report on the design and implementation of an evaporation blocker for cone-plate and plate-plate geometries. In addition to minimizing sample evaporation by trapping the saturated vapor inside a sample chamber, an important feature of the evaporation blocker is the suppression of solvent transport through condensation. Validation tests with DI-water, hydrogels, and colloidal suspensions demonstrate that the new accessory reduces solvent evaporation significantly more than commercially available environmental control chambers. Experiments were also performed to show that the evaporation blocker enables a variety of experiments on thermoresponsive complex fluids that were previously impossible, and provide new insight into the molecular organization of these materials. The evaporation blocker was specifically designed for an Anton Paar MCR rheometer, but the fundamental principles are widely applicable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.