Abstract
The electron beam energy monitoring device, previously developed and tested under standard laboratory conditions using electron beams in the energy range 4–12 MeV, has now been tested under industrial irradiation conditions in high-energy, high-power electron beam facilities. The measuring instrument was improved in order to measure high peak current delivered at low pulse repetition rate as well. Tests, with good results, were carried out at two different EB plants: one equipped with a LUE-8 linear electron accelerator of 7 MeV maximum energy used for cross-linking of cables and for medical device sterilization, and the other with a 10 MeV Rhodotron type TT 100 used for in-house sterilization.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.