Abstract

Proton imaging has been proposed as an alternative low-dose imaging technique for acquiring relative stopping power (RSP) maps and for patient positioning. In proton therapy and hadrontherapy, the water equivalent thickness (WET) of a material represents its radiological thickness, however, its calculation requires computationally intensive methods or approximate solutions. At IEM-CSIC, we have developed a scanner prototype for imaging with protons composed of two double-sided silicon strip detectors (DSSSD) and fast scintillators with high energy resolution, detectors that are commonly used in Experimental Nuclear Physics. Two custom-made samples of aluminum, air, and polymethyl methacrylate (PMMA) were imaged and their proton radiography images (pRads) are presented in terms of WET. Pattern reproduction, spatial resolution, and sensitivity to di↵erent materials were studied. We present pRads of our samples obtained using a rather traditional approach to compute WET values. All three materials of the studied samples, aluminum, PMMA, and air, were well distinguished; the resulting WET values were in good agreement between samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.