Abstract

To evaluate an adjustable mandibular positioning appliance for treatment of snoring and sleep apnea. One hundred thirty-four patients with baseline apnea/hypopnea index (AHI) of 37 +/- 28 events/h (mean +/- SD) received the appliance. The efficacy of the appliance was assessed by the following investigations, performed at baseline and with the appliance: polysomnography, Epworth sleepiness scale, bedpartners' assessment of snoring severity, patients' assessment of side effects, and overall satisfaction. Thirteen patients were lost to follow-up. An additional 46 patients had no follow-up polysomnography, but answered the questionnaires. A total of 75 patients had polysomnography at baseline and with the appliance. We found a significant reduction in AHI from 44 +/- 28 events/h to 12 +/- 15 events/h (p < 0.0005) and a reduction in the arousal index from 37 +/- 27 events/h to 16 +/- 13 events/h (p < 0.05). Epworth scores fell from 11 +/- 5 to 7 +/- 3 (p < 0.0005). Bedpartners' assessment revealed marked improvement in snoring. For example, at baseline 96% of patients were judged to snore loudly "often" or "always" by their bedpartners, whereas only 2% were judged so while using dental appliance. The most frequent side effect was teeth discomfort, present "sometimes" or "often" in up to 32% of patients. Follow-up clinical assessment in 121 patients conducted on the average 350 days after the insertion of the appliance revealed that 86% of patients continued to use the appliance nightly; 60% were very satisfied with the appliance, 27% were moderately satisfied, 11% were moderately dissatisfied, and 2% were very dissatisfied. We conclude that the adjustable mandibular positioning appliance is an effective treatment alternative for some patients with snoring and sleep apnea.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call