Abstract

Purpose Verify if the use of thyroid shield was effective in reducing the radiation dose received by the thyroid in mammography Materials and methods Using a General Electric Senograph DMR mammographer with two bucky s thermoluminescent dosimeters and two phantoms one of body and the other of breast were made measurements in cranio caudal and medio lateral oblique views The variation was made with and without thyroid shield and in the breast without shield on both views Results After collecting and analyzing data it was verified a reduction of the entrance skin doses on thyroid with protection Conclusion Given the significant reduction of the dose and the principle ALARA we can conclude that the shield should be used by the Radiology service However the use of the shield may provide artifacts making difficult the analysis of the breast tissue and leading to the examination repetition increasing the radiation from the mammary gland as well as the other organs

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.