Abstract

Currently used methodologies for quality control of residual leukocytes and erythrocytes in leukocyte-depleted plasma are either expensive or time-consuming. It has been proposed that urine dipsticks could be used as a screening method for residual erythrocytes. The aim was, therefore, to evaluate if urine dipsticks could be used to detect residual erythrocytes and also residual leukocytes in leukocyte-depleted plasma. Dilution series ranging over the decision limits for residual erythrocytes and leukocytes were prepared. Positive, negative and overall agreements, as well as the precision and joint frequency distributions, were calculated for five dipstick analyzers and their corresponding dipsticks. Twenty-four consecutive leukocyte-depleted donor plasma samples were also tested. None of the dipstick analyzers had both a high positive and a high negative agreement. Accordingly, none of the analyzers were able to discriminate between cell concentrations close to the decision limits. The inconsistency count revealed differences in precision between the dipstick analyzers. In the 24 consecutive donor samples, no significant correlation between the dipstick analyzers and the reference methods were found. In conclusion, urine dipsticks are not suitable for quality control of residual leukocytes and erythrocytes in leukocyte-depleted donor plasma.