Abstract

A liquid micro-flow calibration rig is developed, which is consisted of a syringe pump and a gravimetric weighing tank system. Calibration flow rate range is from 0.001 mL/min~10 mL/min. The weighing tank system has been improved to reduce influence of evaporation. The syringe pump was calibrated using the weighing tank to check an influence of evaporation error. The calibrated syringe pulse factors all agreed within ±0.1%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.