Abstract

Knowing the uncertainty associated with an analytical method is necessary to evaluating the quality of the results. In this work, the uncertainty associated with the analytical measurement of polycyclic aromatic compounds (PAHs) in ambient air using thermal desorption coupled with gas chromatography-mass spectrometry (TD-GC/MS) was evaluated. The study focused on analyses of the particulate phase (PM10), and the results were compared to those obtained with a conventional solvent extraction (Soxhlet-GC/MS). The main sources of the combined uncertainty were the calibration curve, recovery of the extraction stage, area or volume of the sample and volume of air. TD-GC/MS reported a lower combined uncertainty (< 10%) than Soxhlet-GC/MS (values between 11% and 18%) for the measurement of particle-bound PAHs in the air, which was primarily contributed by the recovery of the analyzed PAHs and the calibration curves.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.