Abstract

The aim of this study was to evaluate the mechanical behavior of different rigid fixation methods in mandible angle fractures. Three different plates were tested: one 4-hole grid miniplate, one 8-hole curved grid miniplate, and one 4-hole straight miniplate.For the loading tests, 30 polyurethane hemimandibles sectioned at the angle area to simulate a simple angle fracture were used: 10 hemimandibles for each group. The mechanical evaluation revealed that the straight plate positioned at the tension zone presented the best results in supporting vertical loading in the predefined displacement moments of 1, 3, and 5 mm. Regarding the grid plates, there was no statistically significant difference between the smaller and the larger plate.According to this sample, the straight plate positioned in the tension zone presented better results than the other 2 different grid plates positioned at the neutral zone.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.