Abstract
In the current work, we focused on the evaluation of thermal neutron doses in mixed gamma/thermal neutron field using borated Fricke gel with 23 ppm of 10B isotope. The dose response using two measuring protocols and the optical density stability of borated-Fricke gel (BFG) dosimeter were studied for low dose rate of thermal neutron beam from Am-Be source. The performance parameters of BFG were calculated where the critical level and the detection limit were found to be 7 and 13 mGy, respectively. The radiation yield for BFG was found to be 338 ions per 100 eV of thermal neutron dose. This kind of gels may help in the dosimetry of thermal neutrons used in boron neutron capture therapy technique and also in the industrial applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.