Abstract

Acetabular hip joint components manufactured from gamma-sterilized ultra high molecular weight polyethylene (UHMWPE), gamma cross-linked UHMWPE, or polycarbonate-urethane (PCU) polymers were evaluated in a hip joint simulator, using cobalt alloy femoral components, for at least 5 million cycles. The volume of material losses due to wear was calculated for each type of sample, based upon mass loss measurements, every 500,000 cycles. The loss of material for the conventional UHMWPE was much higher than for the cross-linked UHMWPE, showing about a 70% reduction in wear due to cross-linking. The material loss for the PCU samples appears to have been at least 24% lower than for the cross-linked UHMWPE. Based upon these results, the PCU material seems to have potential for use as an alternative bearing material to UHMWPE for total hip replacement surgeries.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.