Abstract

In this work the peak kilovoltage (kVp), practical peak voltage (PPV) and air kerma rate were measured with the noninvasive meters Radcal Accu-kV® Diagnostic Sensor™ model 40×12-W, and PTW Diavolt. The results were compared in order to ensure the quality control, compare the meters and establish the new quantity PPV, at the Calibration Laboratory of IPEN. These tests were performed using the standard diagnostic radiology quality beam RQR5, and the results are in good agreement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.