Abstract

There is limited literature examining the accuracy of the VITEK 2 Advanced Expert System (AES) in characterisation of β-lactamase resistance patterns. We present a prospective single centre study to better ascertain the performance characteristics of this program. The VITEK 2 AES interpretation was compared to established laboratory phenotypic methods. The overall sensitivity for detection of broad-spectrum β-lactamase by the AES was 95%, with a specificity of 78%. One or more discrepancies were noted in 36% of samples, with the majority of these (87/100) due to incorrect 'overcall' of a resistance mechanism. AES characterisation of AmpC resistance mechanisms was excellent. In contrast, the AES had poor specificity in classifying extended spectrum β-lactamases (ESBLs). As a screening aid, the AES can be a valuable tool. However, optimal use requires an adequate working knowledge of resistance mechanisms in order to correctly interpret and accept the result output.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.