Abstract

Haemoglobin screening methods need to be highly sensitive to detect both low and high haemoglobin levels and avoid unnecessary rejection of potential blood donors. The aim of this study was to evaluate the accuracy of measurements by HemoCue in blood donors. Three hundred and fourteen randomly selected, prospective blood donors were studied. Single fingerstick blood samples were obtained to determine the donors' haemoglobin levels by HemoCue, while venous blood samples were drawn for measurement of the haemoglobin level by both HemoCue and an automated haematology analyser as the reference method. The sensitivity, specificity, predictive values and correlation between the reference method and HemoCue were assessed. Cases with a haemoglobin concentration in the range of 12.5-17.9 g/dL were accepted for blood donation. Analysis of paired results showed that haemoglobin levels measured by HemoCue were higher than those measured by the reference method. There was a significant correlation between the reference method and HemoCue for haemoglobin levels less than 12.5 g/dL. The correlation was less strong for increasing haemoglobin levels. Linear correlation was poor for haemoglobin levels over 18 g/dL. Thirteen percent of donors, who had haemoglobin levels close to the upper limit, were unnecessarily rejected. HemoCue is suitable for screening for anaemia in blood donors. Most donors at Yazd are males and a significant percentage of them have haemoglobin values close to the upper limit for acceptance as a blood donor; since these subjects could be unnecessarily rejected on the basis of HemoCue results and testing with this method is expensive, it is recommended that qualitative methods are used for primary screening and accurate quantitative methods used in clinically suspicious cases or when qualitative methods fail.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call