Abstract
Current guideline recommends the use of two identification methods for Neisseria gonorrhoeae. Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF) is now used for primary identification and may be sufficient for definitive identification of N. gonorrhoeae. The performance of three secondary tests (BactiCard, RapID NH and NET test) were compared using 45 bacterial isolates, including 37 Neisseria species. These secondary tests demonstrated diminished specificity (67% - 88%) for N. gonorrhoeae compared with MALDI-TOF. Additionally, data from six clinical microbiology laboratories was used to compare confirmatory test costs and the agreement of results with MALDI-TOF. Discrepancies were documented for 9.4% of isolates, though all isolates (n= 288) identified by MALDI-TOF as N. gonorrhoeae were confirmed by the reference laboratory. These data demonstrate that MALDI-TOF alone is sufficient for N. gonorrhoeae identification, as secondary did not add diagnostic value but do add costs to the testing process.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.