Abstract
The aim of this study was to evaluate a new type of assay for the phenotypic detection of methicillin-resistant Staphylococcus aureus (MRSA). The assay is based on a point-of-care compatible two-photon excitation fluorescence detection technology (TPX). A collection of 243 epidemic MRSA isolates was tested in addition to 138 sporadic MRSA and 101 negative control strains. The assay proved to be both sensitive (97.9%) and specific (94.1%) in the identification of MRSA, with adequate positive (98.4%) and negative (92.2%) predictive values. The time required for obtaining a positive test result was less than 14 h for 99.0% of the MRSA true-positive samples. After a test run, the selectively enriched reaction mixtures may be recovered and further studied by molecular or standard phenotypic methods. The main benefits of the TPX methodology include a simple assay procedure, low reagent consumption, and a high-throughput capacity.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: European Journal of Clinical Microbiology & Infectious Diseases
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.